*If the Devices status is OK - just PWR LED is lit, but ethernet connection was not correctly established (no LED is lit on left side), you can try to reconnect the ethernet cable without power cycling the device and check the network status LEDs.
*If the Devices status is OK - just PWR LED is lit, but ethernet connection was not correctly established (no LED is lit on left side), you can try to reconnect the ethernet cable without power cycling the device and check the network status LEDs.


==== Network connection init ====
==== Next steps ====
*Now, device is ready to start communication with Pixet or to change network settings.
*Now, device is ready to start communication with Pixet or to change network settings.
== 4. Network settings ==
== 4. Network settings ==

Revision as of 12:33, 20 December 2023

Preliminary version.

1. Before you start

  • Check the contents: WidePIX POE device, +24V DC Adapter (optional), Ethernet cable, Quick couplings, memory stick with the software.
  • Warning: the sensor is very fragile – Avoid any mechanical shock!
  • Don’t operate device without active water cooling!

2. Device connections

Using the POE injector

  • Connect the ethernet cable between detector and your network
  • Connect the ethernet cable thru the POE injector (Class 6)

Using DC Adapter

  • Connect the ethernet cable between detector and your network
  • Connect the +24V DC adapter to the device

3. Device power up

Connector panel LEDs

  • There 2 types of Status LEDs: Network status and Device status
  • Network status LEDs are placed on the left side of the connector board (LINK-1 and LINK-2)
Network status LEDs
LINK-1 LINK-2
1000Mbps, no activity Off On
1000Mbps, activity Off Blink
100Mbps, no activity On Off
100Mbps, activity Blink Off
10Mbps, no activity On On
10Mbps, activity Blink Blink
No link Off Off
  • Device status LEDs are placed on the right side of the connector board (PWR, ERR, STAT-1, STAT-2, STAT-3)

Device init

  • When proper power supply is connected (POE or adapter), device should power up. It is normal that each LED slightly blink during device's initialization, after some time (1 - 3 seconds) initialization is done and just PWR LED is lit on the right side.
  • In case of wrong initialization ERR LED starts to blink (in the number of blinks in one period is coded device's error state, shoot a 5 seconds video for tech support guy). Try to power cycle the device.

Network connection init

  • During the Device init process network connection is established. When connection is correctly established LINK-2 LED should irregullary blink (device will not properrly work with other type of link than 1000Mbps!).
  • If the Devices status is OK - just PWR LED is lit, but ethernet connection was not correctly established (no LED is lit on left side), you can try to reconnect the ethernet cable without power cycling the device and check the network status LEDs.

Next steps

  • Now, device is ready to start communication with Pixet or to change network settings.

4. Network settings

  • Device is by default configured to use with DHCP
  • Other IPv4 methods which device can use are:
    • Auto IP (169.254.x.y)
    • Static IP (192.168.1.100)

IPv4 method change

  • Device must be power cycled before IPv4 method change.
  • Attach magnet to the magnetic sensor (magnetic sensor is placed on connector board just under labels STAT-2 and STAT-3), when the magnetic field is recognised, LED STAT-1 is lit. You have just 5 seconds from device's power up to attach the magnet! Otherwise IP address mode change is dissabled to prevent unwanted IP address changes. When dissabled LED STAT-1 blinks fast (power cycle needed).
  • Hold the magnet in front of the magnetic sensor for approximately 8 - 10 s, after that, IPv4 method change is enabled, which is indicated with illumination of both STAT-2 and STAT-3 LEDs.
  • Movement of the magnet in front of the magnetic sensor leads to the change of the state of STAT-2 and STAT-3 LEDs.
  • Each LED state is connected to various IP settings.
  • There are 3 IPv4 methods: Auto IP (STAT-3 LED is lit), DHCP (STAT-2 LED is lit) and Static IP (both STAT-2 and STAT-3 LEDs are lit).
  • To change the method, just move with magnet until the needed combination of LEDs is achieved. Then move magnet away from magnetic sensor and wait approximately 8 - 10 seconds until both STAT-2 and STAT-3 LEDs start to blink, which means setting the new IPv4 method.
  • When LEDs stop to blink, new IPv4 method is set.
  • Power cycle the device.